Titre : | Effect of audiological testing on programmable ventriculoperitoneal shunts (2022) |
Auteurs : | Ayeshah Abdul-Hamida ; Gulam Zilani ; Catriona Bryant ; A. Bryant ; Ali Qureishi ; Sanjeeva Jeyaretna |
Type de document : | Article |
Dans : | International Journal of Audiology IJA (Vol. 61, n°2, Février 2022) |
Article en page(s) : | p. 163-169 |
Note générale : | https://doi.org/10.1080/14992027.2021.1881174 |
Langues: | Anglais |
Descripteurs : |
HE Vinci Bloc auriculo ventriculaire ; Dépistage auditif du nouveau-né ; Test auditifAutres descripteurs Casque audio ; Equipement audiometrique |
Résumé : |
Objective
This study aimed at measuring the magnetic field strength of commonly used types of audiological testing equipment and determine their effects on the three most commonly used programmable ventriculoperitoneal (VP) shunts to try and quantify the risk of a VP shunt being reprogrammed during audiological testing. Design In this in vitro study, magnetic field strength was measured for TDH 39 supra-aural earphones, B71 bone vibrator, ALGO 3i probe, Bio-logic Nav Pro probe, Otodynamics otoport insert earphone and Madsen Zodiac tympanometry probe. Study sample Magnetic field strength associated with transducers placed on a model of a skulls having implanted Miethke ProGAV 2.0, Medtronic Strata II and Codman Hakim programmable VP shunts was measured. Results The supra-aural earphones had a magnetic field strength of 14 mT at 0 mm, which dropped to 0 mT at 10 mm away from the transducer. All other equipment had a magnetic field strength of 3.5 mT or less at 0 mm. There was no instance of reprogramming of the shunts by the transducers. Conclusions The findings suggest that the risk of inadvertent valve-reprogramming by the transducers is extremely small. However, care should be taken to avoid placing any of the transducers directly over the shunt. |
Accès : | Contactez la bibliothèque d'Ixelles si le lien vers la ressource électronique ne fonctionne pas |
Disponible en ligne : | Oui |
En ligne : | https://login.ezproxy.vinci.be/login?url=https://www.tandfonline.com/doi/full/10.1080/14992027.2021.1881174 |